Print batch worklist, Print report, Recalc – Luminex IS Version 2.3 User Manual
Page 42: Replay batch, Report raw fluorescence, Resume, Sample probe down, Sanitize, Save and load, Save only
![background image](/manuals/336977/42/background.png)
Luminex IS Software Manual for Version 2.3
x
MAP Technology
32
PN 89-00002-00-110 Rev. B
Print Batch Worklist
Prints the status of each of the commands in the command list, as
well as the microtiter plate graphic.
Print Report
Prints the data interpretation report including the standards, controls,
graph of standards, and Samples data.
Recalc
Recalculates an affected test after a change is made to lot
information. There are two options: Manual recalculate and auto. If
the auto checkbox is selected, the Recalc button (manually
recalculate) is grayed out. In the auto selection mode, changes
automatically trigger the system to reanalyze the affected test. If you
are not using the auto mode, the ReCalc button is available after you
make a change such as changing a formula or lot, or invalidating or
validating a standard or control.
Replay Batch
Replays batch sample acquisition. All batch commands are replayed
allowing for batch acquisition viewing exactly as in real-time. You
cannot replay New Advanced Batches. When running a Replay
Batch command, the original acquisition data is not altered. A new
file is created. The most common use is for system demonstration
and training.
Report Raw Fluorescence
Enables the median fluorescence intensity (MFI) to appear on the
Analyte Report.
Resume
Resumes the process that was paused.
Sample Probe Down
Raises and lowers the sample probe to adjust the probe height. This
adjustment is recommended when using different microtiter plates or
when changing sample probes. Refer to the Luminex 200 System
Manual or the Luminex 100 IS User Manual for Version 2.3 for the
adjusting procedure.
Sanitize
Uses the Luminex XYP reservoir due to volume requirement. The
sanitize command performs a similar function as the alcohol flush
command. However, the sanitize command uses 10% to 20%
household bleach to decontaminate sample lines and the cuvette after
biohazard contact.
Save and Load
If selected, displays the run batch tab immediately after you set up
batch information and click Finish.
Save Only
Allows you to save the batch information. User can open the batch
later to acquire data.