heater block temperature is out of range or experiencing
unacceptable temperature conditions, including these items:
•
Luminex XYP heater block temperature time-out
•
heat circuit failure
•
temperature out-of-range sensing
•
temperature change since calibration, or a change in channel
temperature
Table 1 defines the values listed in the System Monitor. These values
are useful for diagnostic purposes when communicating with
Luminex Technical Support.
Table 1. System Monitor Values
Property
Value Units of Measure
Air Pressure
PSI, air pressure to the sheath
container from the air pump
Sheath Pressure
PSI, sheath pressure from the
sheath container through the
system
Delta Cal Temperature
°C, temperature deviation from
the last calibration
Board Temperature
°C, temperature of the analog
board
DD Temperature
°C, DD temperature at the U
block inside the optics platform
CL1 Temperature
°C, CL1 temperature at the U
block inside the optics platform
CL2 Temperature
°C, CL2 temperature at the U
block inside the optics platform
XYP Board Temperature
°C, temperature of the XYP
board inside the Luminex XYP
instrument
XYP Heater Temperature
°C, temperature of the XYP
heaters inside the Luminex
XYP instrument
XYP Heater Temperature In
Range
Indicates if the XYP heater is in
the set range