Luminex IS Version 2.3 User Manual
Page 34

Luminex IS Software Manual for Version 2.3
x
MAP Technology
24
PN 89-00002-00-110 Rev. B
high or low end. Examples of out-of-range sample labeling for non-
competitive assays are shown in Table 6. Figure 18 shows what a
standard curve for a non-competitive assay should look like.
Figure 18 Non-Competitive Assay
Examples of out-of-range sample labeling for competitive assays are
shown in Table 7. Figure 19 shows what a standard curve for a
competitive assay should look like.
Table 6. Non-Competitive Out-Of-Range Labeling
Condition
Concentration Label
MFI of Standard Referenced
Left of Curve
< min concentration standard
Lowest
Right of Curve
> max concentration standard
Highest
Table 7. Competitive Out-Of-Range Labeling
Condition
Concentration Label
MFI of Standard Referenced
Left of Curve
< min concentration standard
Highest
Right of Curve
> max concentration standard
Lowest