4. Replace the sample probe and tightly screw in the Cheminert fitting.
5. Use the software to perform an automatic probe height adjustment.
NOTE: Perform an automatic probe height adjustment any time the probe
is .
FIGURE 15.
The Sample Probe
Calibrating the FLEXMAP 3D System
Calibrate the FLEXMAP 3D system weekly as part of regularly scheduled maintenance.
There are several different ways to calibrate the system using the software.
Visually Inspect the FLEXMAP 3D Instrument
Make sure the instrument is idle, so there are no moving parts. Open all of the FLEXMAP 3D
instrument doors and visually inspect for leaks, corrosion, and other signs of improper
function. Check all visible tubing connections.
Monthly Maintenance
Clean the exterior surfaces monthly.
To clean exterior surfaces:
Maintenance and Cleaning