Plumbing and connection options

A.8 Option 4 plumbing and connections

Option 4 uses two sc analyzers with the FILTRAX. The sample from
the FILTRAX goes to the first analyzer which needs to change to a
2-parameter configuration (see

Connect a 2-parameter option,

page 66

). The heated drain hose connects both sc analyzers.The

waste from both analyzers is discharged back into a drain through
the second heated drain hose.

Refer to

Figure 30

and the following instructions for Option 4:

1. Install the FILTRAX into the sample stream. Refer to the

FILTRAX User Manual for more information.

2. Install the first sc analyzer (Analyzer 1):

a. Feed the heated hose from the FILTRAX through the

analyzer opening (

Figure 30

, item 15). Use Seal Plug #1

to secure.

b. Feed the heated drain hose though the analyzer opening

(item 17). Use Seal Plug #1 to secure.

c. Seal the remaining opening with Seal Plug #3.

d. Connect the heated drain power connections. Refer to

section 3.5.5, page 27

.

e. Remove the pre-installed drain tube attached to the valve

block and remove the T-fitting from the drain tube (item 7).
Retain for reuse with Analyzer 2.

f.

Connect the heated drain tube to the valve block connector.

g. Connect the sample line from the FILTRAX to the bottom

inlet on the overflow vessel using the fittings.

h. Change the analyzer to the 2 parameter configuration.

Refer to

Connect a 2-parameter option, page 66

.

i.

Connect one of the sample lines from the heated drain to
the overflow vessel.