5 connect the sample supplies and drain, Allation option, Section 3.4.5 – Hach-Lange PHOSPHAX indoor sc User Manual
Page 23

23
Installation
3.4.5 Connect the sample supplies and drain
Before connecting tubing or cables, determine the option number
that corresponds the system configuration. Refer to
on the option number, determine the sealing plug that will be used
to seal the enclosure openings, refer to
.
When the option number is determined, refer to
Plumbing and connection options on page 65
for installation
information.
Table 1 System configuration options
Location
Filtration
Drain
Number of
analyzers
Sample
lines
(Ch1, Ch2)
Number of
parameters
1
Option
#
Refer to the following section
for more information:
OUTDOOR
Filter Probe sc
Any
1
1
1
1
Filter Probe sc
Heated
1
1
1
2
FILTRAX
Heated
1
1
1
3
FILTRAX
2 heated
2
1
2
4
2 FILTRAX
Heated
1
2
1
5
2 FILTRAX
2 heated
2
2
2
6
INDOOR
Filter Probe sc
Unheated
1
1
1
7
FILTRAX
Unheated
1
1
1
8 a
2
1
2
8 b
2 FILTRAX
Unheated
1
2
1
9 a
2
2
2
9 b
Continuous
sample feed
Unheated
1
1
1
10 a
2
2
2
10 b
2 continuous
sample feeds
Unheated
1
2
1
11 a
2
2
2
11 b
1
For 2-parameter options, refer to
Connect a 2-parameter option on page 66
Table 2 Sealing plug types
Option
Analyzer 1
Analyzer 2
Opening 1
Opening 2
Opening 3
Opening 1
Opening 2
Opening 3
1
Plug 2
Plug 3
Plug 3
—
—
—
2
Plug 2
Plug 1
Plug 3
—
—
—
3
Plug 1
Plug 1
Plug 3
—
—
—
4
Plug 1
Plug 1
Plug 3
Plug 1
Plug 1
Plug 3
5
Plug 1
Plug 1
Plug 1
—
—
—
6
Plug 1
Plug 1
Plug 1
Plug 1
Plug 1
Plug 3
7
Plug 2
Plug 3
Plug 3
—
—
—
8
Plug 1
Plug 3
Plug 3
Plug 3
Plug 3
Plug 3
9
Plug 1
Plug 1
Plug 3
Plug 3
Plug 3
Plug 3
10
Plug 3
Plug 3
Plug 3
Plug 3
Plug 3
Plug 3
11
Plug 3
Plug 3
Plug 3
Plug 3
Plug 3
Plug 3