Fieldbus – Hach-Lange PHOSPHAX indoor sc User Manual
Page 106
106
Fieldbus
8.TH LAST VALUE
40250
Float
2
R
LIST OF VALUES
9.TH LAST VALUE
40252
Float
2
R
LIST OF VALUES
STATUS
MODUL.ERR
40256
Unsigned
Integer
1
R/W
14/10/8/0
configure the level of error for the
state of the modules
STAT.
MODUL.WAR.
40257
Unsigned
Integer
1
R/W
40/30/15
configure the level of warning for the
state of the modules
EXHAUST
CONTROL
40258
Unsigned
Integer
1
R/W
0/1
the analyzer is checking the
exhaust whether it is blocked;
0=OFF, 1=ON
REF
40260
Unsigned
Integer
1
R/W
0/1
to switch the ref measuring on or off
BUBBLE REJECT
40264
Unsigned
Integer
1
R/W
0/1
it's possible to set the function
bubble reject
SEL ADJ METHOD
40271
Unsigned
Integer
1
R/W
0/1
it's possible to set the measuring
interval as dry or wet
ENCLOSU.TEMP.
MIN
40273
Float
2
R
-50/200.0
the minimum temperature inside the
analyzer during the last 24 hours,
intervall starts with power on
ENCLOSU.TEMP.
MAX
40275
Float
2
R
-50/200.0
the maximum temperature inside
the analyzer during the last 24
hours, intervall starts with power on
ERROR LIST
40277
Unsigned
Integer
2
R
" Errors coded bit wise, bit0=TEMP.
< 0 °C/ 32°F?, bit1=ANALYZ. TO
COLD, COOLING FAILED,
HUMIDITY ANALY,
HUMIDITY PROBE, PROBE
MISSING, NO HEAT UP,
CUVSENSOR DEFECT,
TEMPSENS DEFECT, CUVHEAT
DEFECT, CUV TOO HOT, PHOTO
LEVEL LOW,
PHOTO LEVEL HIGH, MODULES
CONTAM., PHOTO LEVEL2 LOW,
PHOTO LEVEL2 HIGH, DRAIN
BLOCKED, SAMPLE1,
bit18=SAMPLE2"
WARNING LIST
40279
Unsigned
Integer
2
R
"Warnings, coded bit wise,
bit0=WARMUP PHASE,
bit1=COOLING DOWN, SERVICE
MODE, REAGENT LEVEL,
CLEAN SOLU LEVEL, ANALYZER
TO COLD, ANALYZER TO WARM,
CUV TOO COOL,
MODULES CONTAM., PHOT
LEVEL LOW, SAMPLE1, PHOT
LEVEL2 LOW, bit12=SAMPLE2"
EDIT NAME
40281
String
8
R/W
Name of LOCATION
(see menu system)
Table 8 Sensor modbus registers
Tag Name
Register #
Data
Type
Length R/W
Discrete
Range
Min/Max
Range
Description