Fieldbus – Hach-Lange PHOSPHAX indoor sc User Manual
Page 102

102
Fieldbus
P2O5 VALUE 2
40056
Float
2
R
measurement value for channel two
as P2O5
DEXT LAST VALUE
40058
Float
2
R
-9.99999/
9.99999
last extinction value
HEATING ON
40060
Unsigned
Integer
1
R/W
0/1/2/3/4/
5/6/7/8/9/
10/11/12
sets the month when the sample
line heating is switched ON;
0=alway OFF, 1=January,
2=Febraury to 12=December
HEATING OFF
40061
Unsigned
Integer
1
R/W
1/2/3/4/5/
6/7/8/9/1
0/11/12
sets the month when the sample
line heating is switched
OFF;1=January, 2=Febraury to
12=December
MEASURE VALUE 2 40062
Float
2
R
actual measurement value from
channel two
EXT MESS 1
40064
Float
2
R
-9.99999/
9.99999
the current measuring extinction
from channel one
EXT MESS 2
40066
Float
2
R
-9.99999/
9.99999
the current measuring extinction
from channel two
EXT REF
40068
Float
2
R
-9.99999/
9.99999
the current reference extinction
ANALYZER
HEATING
40070
Unsigned
Integer
1
R
0/100
the heating of the analyzer
BUSACTION
ACTIVE
40071
Unsigned
Integer
1
R/W
0/1
write a one to this register to start a
bus action (see Field bus control)
BUS ANALY.START
40072
Unsigned
Integer
1
R/W
0/1
you can start the analyzer about the
bus
BUS SERVICE
40073
Unsigned
Integer
1
R/W
0/1
you can start the service mode
about the bus
BUS CLEANING
40074
Unsigned
Integer
1
R/W
0/1
you can start the cleaning mode
about the bus
BUS PREPUMP
REA.
40075
Unsigned
Integer
1
R/W
0/1
you can prepump reagent about the
bus
BUS PREPUMP
CLEA
40076
Unsigned
Integer
1
R/W
0/1
you can prepump cleaning solution
about the bus
BUS PREPUMP
PRO.
40077
Unsigned
Integer
1
R/W
0/1
you can prepump the probe about
the bus
BUS PREPUMP
ALL
40078
Unsigned
Integer
1
R/W
0/1
you can prepump all about the bus
PROBE P. MIN
40079
Float
2
R
0/2.0
integrated value of pressure at
filtration probe, if not yet calculated:
nan
PROBE
PRESSURE
40081
Float
2
R
0/2.0
actual value of pressure at filtration
probe, if not yet calculated: nan
GAIN CORR. 1
40083
Float
2
R/W
0.01/
100.00
Gain correction for channel one
DISCHARGE
CLEAN.
40096
Unsigned
Integer
1
R/W
0/10
discharged values after a cleaning
Table 8 Sensor modbus registers
Tag Name
Register #
Data
Type
Length R/W
Discrete
Range
Min/Max
Range
Description