5 scheduled maintenance, 6 modify from single channel to dual channel, Maintenance – Hach-Lange PHOSPHAX indoor sc User Manual
Page 46

46
Maintenance
2. On the controller TEST/MAINT menu, start a cleaning cycle
with distilled water using the FLUSHING.
3. Clean the canister lid with distilled water.
4. Take the tubing out of the water and start the FLUSHING
function to pump the tubing and the analysis instrument empty.
5. Wipe the canister lids dry and seal the canisters with the
corresponding lids.
6. Remove the canisters and store them in a frost-free place and
in accordance with local regulations.
7. Isolate the system from the mains and the data network.
8. When using a Filter Probe sc, refer to the Filter Probe sc User
Manual for storage information.
9. Install all transport locks.
10. Depending on the duration, remove the system from its
mounting and wrap the system in a protective film or dry cloth.
Store the system in a dry place.
6.5 Scheduled maintenance
lists items that need to be maintained by service personnel
ONLY. Contact the manufacturer for more information.
6.6 Modify from single channel to dual channel
The sc analyzer can be converted from single channel to dual
channel operation and/or continuous sampling. Contact the
manufacturer for more information. Refer to
configuration options.
Table 5 Repair maintenance items
Description
When to replace
Warranty
Reagent pump for sc analyzer (Valve pumps)
According to wear
1 year
Pump head piston pump 10 mL
(Pre-greased cylinder and piston)
1 year
1 year
Switchable Compressor 115/230V
2 years recommended
2 years
Table 6 Conversions
From
To
With
Conversion kit
1-channel operation
2-channel operation
AMTAX sc, PHOSPHAX sc
LZY170
Filter Probe sc
Continuous sampling
AMTAX sc, PHOSPHAX sc
LZY241
Continuous sampling
Filter Probe sc
AMTAX sc, PHOSPHAX sc
LZY242