beautypg.com

Fieldbus – Hach-Lange PHOSPHAX indoor sc User Manual

Page 103

background image

103

Fieldbus

SET OUTMODE
CLE.

40097

Unsigned
Integer

1

R/W

0/1

Set output mode for cleaning;
0=HOLD, 1= TRANSFER VALUE

SET OUTMODE
SER.

40098

Unsigned
Integer

1

R/W

0/1

Set output mode for service mode;
0=HOLD, 1= TRANSFER VALUE

SOFTWARE
PROBE

40099

Float

2

R


0/3.40282
347E+38

the softwareversion of the filtration
probe

STATUS MODULES 40101

Float

2

R

0/100

the state of the modules as float in
percent; nan if not yet calculated

MEAS.UNITS 1

40103

Unsigned
Integer

1

R/W

0/2

Measurement units for channel one
0=mg/l, 2=ppm

APPL.

40104

Float

2

R


0/3.40282
347E+38

the entry is for the application
file;shows the version

TYPE

40106

String

6

R

name of the item/analyzer

STATUS MODULES 40112

Unsigned
Integer

1

R

0/100

the state of the modules as integer
in percent

CLEANING
MODULES

40113

Time2

2

R/W

Date of last filter module cleaning

INTERVAL

40115

Unsigned
Integer

1

R/W


0/1/2/3/4/
5/6/7/8/9/
10/11/12/
13/14/15/
16/17/18/
19/20/21/
22/23

measuring intervall; 0=5 minutes,
1=10minutes,
2=15 minutes to 23=120 minutes,
33=3h, 45=4h, 57=5h, 69=6h,
81=7h, 93=8h, 105=9h, 117=10h,
129=11h, 141=12h, 153=13h,
165=14h, 177=15h, 189=16h,
201=17h, 213=18h, 225=19h,
237=20h, 249=21h, 261=22h,
273=23h, 285=24h

START

40117

Unsigned
Integer

1

R/W


0/1/2/3/4/
5/6/7/8/9/
10/11/12/
13/14/15/
16/17/18/
19/20/21/
22/23

it describes when the analyzer
starts the cleaning (24 hour format)
0=0 o`clock to23=23o´clock

SET INTERVAL

40118

Unsigned
Integer

1

R/W


0/1/3/6/8/
12/24

cleaning intervall ; 0=OFF, 1=1h,
3=3h, 6=6h, 8=8h, 12=12h, 24=24h

NEW MODULES

40119

Time2

2

R

date of the last filter module
exchange

SET PARAMETER 1 40121

Unsigned
Integer

1

R/W


P15/14/1
6

Parameter for channel 1; 15=PO4P,
14=PO4, 16=P2O5

GAIN CORR. 2

40122

Float

2

R/W

0/100.00

Gain correction for channel two

SET PARAMETER 2 40125

Unsigned
Integer

1

R/W


P15/14/1
6

Parameter for channel 2; 15=PO4P,
14=PO4, 16=P2O5

MEAS.UNITS 2

40126

Unsigned
Integer

1

R/W

U0/2

Measurement units for channel two,
0=mg/l, 2=ppm

Table 8 Sensor modbus registers

Tag Name

Register #

Data
Type

Length R/W

Discrete
Range

Min/Max
Range

Description

This manual is related to the following products: