beautypg.com

A.2.1 remove the t-fitting, A.3 drain line considerations, A.4 tubing considerations – Hach-Lange PHOSPHAX indoor sc User Manual

Page 67

background image

67

Plumbing and connection options

A.2.1 Remove the T-fitting

When using the 2-parameter configuration, the T-fitting from the
first analyzer drain tube must be removed and reused to connect
the drain tube from the first analyzer to the second analyzer.

The T-fitting is used to connect the drain tube. To remove the
T-fitting refer to

Figure 26

and the following steps:

1. Remove the drain tube from both ends of the T-fitting.

2. Remove the drain tube assembly.

3. Reconnect the T-fitting as described in Options 4, 6, 8b, 9b,

10b, and 11b.

A.3 Drain line considerations

Always place the drain tubing so that there is a continuous fall
(minimum 3°) and the outlet is clear (not pressurized). Make sure
the drain tubing is no longer than 2 meters.

A.4 Tubing considerations

The analyzer uses four different tubing types for plumbing
connections. The type of tubing used depends on the instrument
configuration:

Ø 3

.

,

2 mm: sample line tubing

Ø 6 mm: unheated drain tubing

Ø 22 mm: heated drain tubing

Ø 32 mm: tubing for the Filter Probe sc

Figure 26 Removing the T-fitting

1

T-fitting to be removed

This manual is related to the following products: