Fieldbus – Hach-Lange PHOSPHAX indoor sc User Manual
Page 105
105
Fieldbus
CLEAN SOLU
LEVEL
40158
Unsigned
Integer
1
R
0/100
cleaninig solution level in percent
AIR FILTER DISPL
40160
Integer
1
R
-32768/
32767
days until next cleaning/exchange
of air filter pads. Negative values
show a cleaning which is overdue
PUMP DISPLAY
40163
Integer
1
R
-32768/
32767
days left until exchanging piston of
pump, negative values show that
exchange is overdue
STRUCTURE
40172
Unsigned
Integer
1
R
0/65535
the entry is for the device driver
file;shows the version
FIRMWARE
40173
Unsigned
Integer
1
R
0/65535
the entry is for the device driver
file;shows the version
LOADER
40174
Float
2
R
0/3.40282
34663852
9E+38
the entry is for the application
file;shows the version of the boot file
OPERATING
HOURS
40176
Unsigned
Integer
2
R
0/999999
99
operating hours of analyzer
PUMP
MEMBR.DISP.
40180
Integer
1
R
-32768/
32767
remaining days for pump membrane
in filter probe
COMPRESSOR
40189
Integer
1
R
-32768/
32767
remaining days for air compressor
LAST
CHA.FACTOR1
40198
Time2
2
R
the date of the last correction factor
for channel one
LAST
CHA.FACTOR2
40200
Time2
2
R
the date of the last correction factor
for channel two
SAMPLE
DETECTION
40211
Unsigned
Integer
1
R/W
0/1/2
output if sample detection detects
low sample amount; 0=Warning,
1=Error, 2=OFF
ACTUAL
MEAS.TIME
40216
Time2
2
R
time of actual measurement value
LAST TIME
40218
Time2
2
R
time of last measurement value
2.ND LAST TIME
40220
Time2
2
R
2.ND LAST TIME
3.RD LAST TIME
40222
Time2
2
R
3.RD LAST TIME
4.TH LAST TIME
40224
Time2
2
R
4.TH LAST TIME
5.TH LAST TIME
40226
Time2
2
R
5.TH LAST TIME
6.TH LAST TIME
40228
Time2
2
R
6.TH LAST TIME
7.TH LAST TIME
40230
Time2
2
R
7.TH LAST TIME
8.TH LAST TIME
40232
Time2
2
R
8.TH LAST TIME
9.TH LAST TIME
40234
Time2
2
R
9.TH LAST TIME
LAST VALUE
40236
Float
2
R
LIST OF VALUES
2.ND LAST VALUE
40238
Float
2
R
LIST OF VALUES
3.RD LAST VALUE
40240
Float
2
R
LIST OF VALUES
4.TH LAST VALUE
40242
Float
2
R
LIST OF VALUES
5.TH LAST VALUE
40244
Float
2
R
LIST OF VALUES
6.TH LAST VALUE
40246
Float
2
R
LIST OF VALUES
7.TH LAST VALUE
40248
Float
2
R
LIST OF VALUES
Table 8 Sensor modbus registers
Tag Name
Register #
Data
Type
Length R/W
Discrete
Range
Min/Max
Range
Description