Maintenance,Troubleshooting
Type 8626, 8712
procedure:
→
To gain access to the stainless steel mesh filter disc, detach the
input flange plate [2] (see „Fig. 21“).
→
Take out the stainless steel mesh filter disc.
→
Clean the stainless steel mesh filter disc [5] using distilled water
(not tap water), acetone, isopropanol or compressed air.
→
Dry the parts after cleaning.
→
Re-insert parts in the correct sequence and position
(see „Fig. 21“). The fine mesh of the filter disc [5] must face
the input flange plate [2].
9.2.2. cleaning and recalibration at the
factory
If the sensor is excessively soiled or damaged by the operating gas,
the device may deviate significantly in terms of the mass flow-rate
measurement. Cleaning or replacement followed by recalibration at
the factory will then be required.
note
• Recalibration must be carried out at the factory as it requires
the use of very precise references and a specific digital
communication system.