Description of the mfm / mfc, Sequence of the steps to be performed, Setting the parameters – Burkert Type 8712 User Manual
Page 24

24
Installationandcommissioning
Type 8006, 8702
7.3.
Description of the mfm / mfc
LEDs
M12 5-pin fixed connector or
Sub-D 9-pin fixed connector
Fieldbus
Sub HD 15-pin fixed
connector Set-point input /
Measured flow-rate output /
Binary inputs / RS232
Screw for earth
connection
8-pin round fixed connector
24 V supply /
Relay outputs
Connection
to the line
Fig. 20: Description of the MFM / MFC
7.4.
sequence of the steps to be
performed
1.
Mechanical installation
2.
Fluid installation
3.
Electrical installation
4.
Set the device parameters
5.
Pressurize the lines with operating fluid
6.
Flush the lines with operating fluid at the calibration pressure
and deaerate completely
7.5.
setting the parameters
7.5.1. setting the bus address
To ensure trouble-free adjustment, reset the device by
switching off the power supply to the device.
The bus address of the device can be set either via the Bürkert
"Mass Flow Communicator" software tool in the "Views" window →
PROFIBUS / DeviceNet / CANopen or directly via the master bus.
The address must be reinitialized after a change on the slave and on
the master. Depending on the bus, it may be necessary to send a
corresponding telegram.