2 trigger source, 3 burst count, 4 internal trigger rate – B&K Precision 4034 - Manual User Manual
Page 60

60
BURSt
Examples:
:TRIG:MODE CONT
:TRIG:MODE BURS
QUERY
Syntax:
:TRIGger:MODE?
Response:
CONT | TRIG | GATE | BURS
4.12.3.2 Trigger Source
This command is used to select the trigger source, for use in the Trigger, Gate and Burst trigger modes.
COMMAND TYPE:
Setting or Query
SETTING
Syntax:
:TRIGger:SOURce
Arguments:
Type:
Character
Options:
MANual – Front panel MAN key
BUS – GPIB trigger (GET or *TRG)
INTernal – Internal trigger
EXTernal – External trigger
Examples:
:TRIG:SOUR BUS
:TRIG:SOUR INT
QUERY
Syntax:
:TRIGger:SOURce?
Response:
MAN | BUS | INT | EXT
4.12.3.3 Burst Count
Used to set the number of cycles to be output in the BURST mode. It is not a standard SCPI command.
COMMAND TYPE:
Setting or Query
SETTING
Syntax:
:TRIGger:BURSt
Arguments:
Type:
NRf
Range:
2 to 999999
Rounding:
To integer value
Examples:
:TRIG:BURS 100
QUERY
Syntax:
:TRIGger:BURSt?
Response:
NRf
Examples:
:TRIG:BURSt?
4.12.3.4 Internal Trigger Rate
Sets the rate of the internal trigger.
COMMAND TYPE:
Setting or Query
SETTING
Syntax:
:TRIGger:TIMer