14 pulse parameter limitations – B&K Precision 4034 - Manual User Manual
Page 32

32
In the SINGLE or DOUBLE pulse mode the instrument defines PERIOD as the time between the 50% points on
the leading edges of two consecutive trigger outputs. DELAY, in double pulse mode, is the time between the leading
edges of the first and second pulse using as a reference point 50% amplitude with fastest transition times.
SETTLING TIME is the interval required for the pulse level to enter and remain in the specified level ACCURACY
RANGE, measured from the 90% AMPLITUDE point.
3.14 Pulse Parameter Limitations
The following formulas express the limits on Period, Width, and Delay.
Single Pulse per Period Modes
(Un-delayed, Delayed, Counted Burst with single pulse mode)
[Period - (Width + Delay)] must be > 10 ns
0.99 * Period must be > (Width + Delay)
Pulse max
= 10.00 s
Pulse min
= (Width + Delay + 10 ns), but not less than 20 ns
Width max
= [(Period * 0.99) - Delay – 10 ns], but not more than 9.89999 s
Width min
= 10 ns
Delay max
= [(Period * 0.99) - Width – 10 ns], but not more than 9.89998 s
Delay min
= 0
Single Pulse Transition Time Restrictions
Width must be > 1.3 * Leading Edge
(Period - Width) must be > 1.3 * Trailing Edge
Double Pulse per Period Modes
(Paired Pulse and Counted Burst with Paired pulses)
Delay must be > Width
0.99 * Delay must be > (Width + 10 ns)
Pulse max
= 10.00 s
Pulse min
= (Width + Delay + 10 ns), but not less than 40 ns
Width max
= [(0.99 * Delay) – 10 ns], but not > 4.85000 s
Width min
= 10 ns