2 high voltage level, 3 low voltage level – B&K Precision 4034 - Manual User Manual
Page 50

50
Range:
0.1Hz to 50MHz
Rounding:
To the resolution of the range.
Examples:
:FREQ 5KHZ
:FREQ 5E3
QUERY
Syntax:
[:SOURce]:FREQuency[:CW|:FIXed]?
Examples:
:FREQ?
Response:
NR3
CONSIDERATIONS: FIXed is an alias for CW.
4.12.1.2 High Voltage Level
This command is used to set the high level of the pulse.
COMMAND TYPE:
Setting or Query
SETTING
Syntax:
[:SOURce]:VOLTage[:LEVel][:IMMediate]:HIGH
Arguments:
Type:
NRf
Units:
MV, V (default)
Range:
-9.5V to +10V
Rounding:
To 10mV
Examples:
VOLT:HIGH 4V
QUERY
Syntax:
[:SOURce]:VOLTage[:LEVel][:IMMediate]:HIGH?
Examples:
:VOLT:HIGH?
Response:
NRf
CONSIDERATIONS:
1)
The high level must be greater than the low level.
2)
The difference between the levels must conform to 0.5V
≤ difference ≤ 10V
3)
The high level may not exceed the high limit.
4.12.1.3 Low Voltage Level
This command is used to set the low level of the pulse.
COMMAND TYPE:
Setting or Query
SETTING
Syntax:
[:SOURce]:VOLTage[:LEVel][:IMMediate]:LOW
Arguments:
Type:
NRf
Units:
MV, V (default)
Range:
-10V to +9.5V
Rounding:
To 10mV
Examples:
:VOLT:LOW 4V
QUERY
Syntax: