19 pulse polarity, 20 channel dependency – B&K Precision 4034 - Manual User Manual
Page 58

58
500ns to 10us
5us to 100us
50us to 1ms
500us to 10ms
4.12.1.19 Pulse Polarity
This command is used to control the polarity of the pulse, which may be normal or complemented. The
COMPement and INVerted parameters are aliases: either may be used.
COMMAND TYPE:
Setting or Query
SETTING
Syntax:
[:SOURce]:PULSe:POLarity
Arguments:
Type:
Character
Options:
NORMal – Normal polarity
COMPlement – complemented
INVerted – complemented
Examples:
:PULS:POL NORM
:PULS:POL INVerted
QUERY
Syntax:
[:SOURce]:PULSe:POLarity?
Examples:
:PULS:POL?
Response:
NORM | COMP
4.12.1.20 Channel Dependency
This command is used to control the dependency of channel 2. It can select either for channel 2 to be
an independent channel, or set it to be dependent on channel 1. This means that channel 2 will have the
same clock and trigger, as well as same frequency and period as channel 1.
COMMAND TYPE:
Setting or Query
SETTING
Syntax:
:SOURce:COUP
Arguments:
Type:
Character
Options:
ON or 1– Dependent on channel 1
OFF or 0 – Independent
Examples:
:SOUR:COUP 1
:SOUR:COUP OFF
QUERY
Syntax:
:SOURce: COUP?
Examples:
:SOUR:COUP?
Response:
1 | 0