Frequency, Level, Signal acquisition – Atec Rohde-Schwarz_FSL Series User Manual
Page 35
Version 09.00, December 2008
R&S
®
FSL Spectrum Analyzer
35
R&S
®
FSL-K93 WiMAX™/WiBro IEEE 802.16 OFDMA SISO
analysis (IEEE 802.16-2004, IEEE 802.16-2004/Cor1-2005,
IEEE 802.16e-2005, P802.16-Rev2/D3 WiMAX™ and WiBro)
The specifications of the R&S
®
FSL-K93 option are based on the data sheet of the R&S
®
FSL spectrum analyzer.
Specifications apply under the following conditions: 30 minutes warm-up time at ambient temperature, specified environmental
conditions met, calibration cycle adhered to, and all internal automatic adjustments performed. "Typical values" are designated with
the abbreviation "typ." These values are verified during the final test but are not assured by Rohde & Schwarz. "Nominal values" are
design parameters that are not assured by Rohde & Schwarz. These values are verified during product development but are not
specifically tested during production.
Frequency
RF input
R&S
®
FSL3 50
MHz
17
to 3 GHz
R&S
®
FSL6 50
MHz
17
to 6 GHz
Frequency range
R&S
®
FSL18 50
MHz
17
to 18 GHz
model
serial number or lower
R&S
®
FSL3
18
100703
R&S
®
FSL6
18
100604
R&S
®
FSL3 100945
R&S
®
FSL6 100759
1.44 MHz to 20 MHz
model
serial number or higher
R&S
®
FSL3
18
100704
R&S
®
FSL6
18
100605
R&S
®
FSL3 100946
R&S
®
FSL6 100760
1.44 MHz to 32 MHz
Sampling rate F
s
R&S
®
FSL18
1.44 MHz to 32 MHz
Level
Level range
RF input
60 dBm to +30 dBm
auto
Level setting
manual
Signal acquisition
IEEE 802.16-2004 OFDMA SISO
IEEE 802.16-2004/Cor1-2005 OFDMA
SISO
IEEE 802.16e-2005 OFDMA SISO
P802.16-Rev2/D3 OFDMA SISO
Supported standards
IEEE 802.16e-2005 based WiBro
Capture length
24 µs to 15.6 ms
Gate length
24 µs to capture length
Number of analyzed subframes
1 to 10922 subframes
EVM versus symbol, burst summary list,
constellation versus symbol, bit stream
capture length
19
FFT spectrum, CCDF
capture length or gate length
Result length
result summary, power versus time, EVM
versus carrier, spectrum flatness,
spectrum flatness difference, group delay,
frequency error versus sample, phase
error versus sample
capture length
19
or
1 to 10922 subframes
Frame length
2 ms to 15 ms
spectrum mask
2.5 ms to 16000 s, auto
Sweep time
adjacent channel power ratio (ACPR)
10 ms to 16000 s
Sweep count
1 to 32767
Trigger modes
RF input
free run, external, power
17
5 MHz to 50 MHz with restricted functionality depending on bandwidth (power trigger, auto level, IF overload).
18
With tracking generator.
19