Result display, Adjustable parameters, Measurement uncertainty (nominal) – Atec Rohde-Schwarz_FSL Series User Manual
Page 31
Version 09.00, December 2008
R&S
®
FSL Spectrum Analyzer
31
Result display
min./mean/max. values
peak vector error
Result list
min./mean/max. values
burst EVM
I/Q offset
gain imbalance
quadrature error
center frequency error
chip clock error
rise time
fall time
mean burst power
peak burst power
crest factor
Power versus time
up ramp/down ramp
EVM
EVM versus symbol
Error versus preamble
frequency error versus preamble
phase error versus preamble
Spectrum
spectrum mask, ACPR, spectrum FFT
Constellation
constellation
diagram
Statistics
bit
stream
PLCP header
CCDF
Limit check
values in line with standard
result list, power versus time, EVM,
spectrum mask, ACP
Adjustable parameters
Tracking
phase
ON/OFF
timing ON/OFF
level ON/OFF
Measurement uncertainty (nominal)
level
23 dBm to +15 dBm
average of 20 bursts, 11 Mbps CCK with short PLCP, burst EVM
Residual EVM
f = 2.442 GHz
1.8 %
Frequency error
Lock range
±0.6 MHz
Uncertainty
1 Hz + reference frequency uncertainty
test of spectrum mask
0.2 dB
output power
f < 3 GHz
0.5 dB
3 GHz
f 6 GHz
0.8 dB
Level uncertainty
ACPR
0.5 dB