beautypg.com

General specifications – Atec Rohde-Schwarz_FSL Series User Manual

Page 11

background image

Version 09.00, December 2008

R&S

®

FSL Spectrum Analyzer

11

General specifications

Remote control
LAN interface

10/100BaseT, RJ-45

IEC/IEEE bus (GPIB)

R&S

®

FSL-B10

SCPI 1997.0

Display
Resolution

640 × 480 pixels

Pixel failure rate

<2 × 10

–5

Mass memory
Mass memory

flash disk (internal), USB memory stick
(not supplied)

Data storage

>500 instrument settings and traces

Temperature

operating temperature range

+0 °C to +50 °C

permissible temperature range

+0 °C to +55 °C

storage temperature range

–40 °C to +70 °C

Climatic loading

+25 °C/+40 °C at 85 % relative humidity
(IEC 60068-2-30)

Mechanical resistance

sinusoidal

IEC 60068-2-6

Vibration

random

IEC 60068-2-64

Shock

40 g shock spectrum,
in line with MIL-STD-810E, method 516.4
procedure 1, IEC 60068-2-27

Power supply
Input voltage range, AC, nominal

100 V to 240 V

AC supply frequency

50 Hz to 400 Hz

Input current, AC

0.9 A to 0.3 A

Input voltage range, DC, nominal

R&S

®

FSL-B30

10 V to 28 V

Input current, DC

R&S

®

FSL-B30

8.0 A to 2.2 A

Power consumption

typ. 45 W, max. 65 W with all options

Safety

IEC 61010-1, EN 61010-1, UL 61010B-1,
CSA C22.2 No. 1010-1

Test mark

VDE, GS, CSA, CSA-NRTL

EMC

in line with European EMC Directive
89/336/EEC and the new EMC Directive
2004/108/EC
including:

- IEC/EN 61326 class B (emission)
- CISPR 11/EN 55011/group 1
class B (emission)

- IEC/EN 61326 Table A.1 (immunity,
industrial)

with handle

408.8 mm × 158.1 mm × 465.3 mm
(16.09 in × 6.22 in × 18.32 in)

Dimensions (W × H × D)

without handle

342.3 mm × 158.1 mm × 367.0 mm
(13.48 in × 6.22 in × 14.45 in)

without options

<7 kg (<15.43 lb)

Weight

with battery pack

<8 kg (<17.64 lb)

Recommended calibration interval

1

year

operation with external reference

2 years