Corrected system performance

The specifications in this section
apply for measurements made
with the Agilent E836xA PNA
Series microwave network
analyzer with the following
conditions:
• 10-Hz IF bandwidth
• no averaging applied to data
• environmental temperature of

25°C ±5°C, with less than 1°C
deviation from the calibration
temperature

• isolation calibration with an

averaging factor of 8

1. The system dynamic range is calculated as the

difference between the noise floor and the source
maximum output power. The effective dynamic
range must take measurement uncertainties and
interfering signals into account.

2. The test port system dynamic range is calculated

as the difference between the test port noise
floor and the source maximum output power. The
effective dynamic range must take measurement
uncertainties and interfering signals into account.

3. The direct receiver access input system dynamic

range is calculated as the difference between the
direct receiver access input noise floor and the
source maximum output power. The effective
dynamic range must take measurement uncertainties
and interfering signals into account. This set-up
should only be used when the receiver input will
never exceed its damage level. When the analyzer
is in segment sweep mode, the analyzer can have
pre-defined frequency segments which will output
a higher power level when the extended dynamic
range is required (i.e. devices with high insertion
loss), and reduced power when receiver damage
may occur (i.e. devices with low insertion loss).
The extended range is only available in one-path
transmission measurements.

4. May be limited to 100 dB at particular frequencies

below 500 MHz due to spurious receiver residuals.
Methods are available to regain the full dynamic
range.

System dynamic range

1

Description

Specification (dB) Typical (dB) at direct

at test port

2

receiver access input

3

Dynamic range

Standard configuration and standard power range (E836xA)

45 to 500 MHz

4

94

N/A

500 MHz to 2 GHz

119

N/A

2 to 10 GHz

122

N/A

10 to 20 GHz

123

N/A

20 to 30 GHz

114

N/A

30 to 40 GHz

110

N/A

40 to 45 GHz

109

N/A

45 to 50 GHz

104

N/A

Extended configuration and standard power range (E836xA-Option 014)

45 to 500 MHz

4

94

132

500 MHz to 2 GHz

119

138

2 to 10 GHz

122

137

10 to 20 GHz

121

136

20 to 30 GHz

111

123

30 to 40 GHz

107

119

40 to 45 GHz

105

116

45 to 50 GHz

100

111

Standard configuration and extended power range and bias-tees (E836xA-Option UNL)

45 to 500 MHz

4

92

N/A

500 MHz to 2 GHz

117

N/A

2 to 10 GHz

120

N/A

10 to 20 GHz

121

N/A

20 to 30 GHz

112

N/A

30 to 40 GHz

108

N/A

40 to 45 GHz

105

N/A

45 to 50 GHz

99

N/A

Extended configuration and extended power range and bias-tees

(E836xA-Option UNL and Option 014)

45 to 500 MHz

4

92

130

500 MHz to 2 GHz

117

136

2 to 10 GHz

120

135

10 to 20 GHz

119

134

20 to 30 GHz

109

121

30 to 40 GHz

105

117

40 to 45 GHz

101

112

45 to 50 GHz

95

108