Measurement throughput summary

Cycle time vs. IF bandwidth

1

Cycle time vs. number of points

1

Instrument state: preset condition, 201 points,

Instrument state: preset condition, 35 kHz IF

CF = 1 GHz, Span = 100 MHz, correction off. Add

bandwidth, CF = 1 GHz, Span = 100 MHz, correction

21 ms for display on. Cycle time includes sweep and off. Add 21 ms for display on. Cycle time includes
re-trace time.

sweep and re-trace time.

Cycle time

1

,

2

(ms)

Number of points

51

201

401

1601

Start (fundamental band), IFBW = 35 kHz

Uncorrected and one-port cal

21

23

28

65

Two-port cal

52

57

70

152

Start 45 MHz kHz, stop 10 GHz, IFBW = 35 kHz

Uncorrected and one-port cal

71

79

84

110

Two-port cal

153

171

182

243

Start 45 MHz, stop 20 GHz, IFBW = 35 kHz

Uncorrected and one-port cal

103

116

121

139

Two-port cal

216

245

256

303

Start 45 MHz, stop 40 GHz, IFBW = 35 kHz

Uncorrected and one-port cal

145

181

190

232

Two-port cal

293

367

382

428

Start 45 MHz, stop 50 GHz, IFBW = 35 kHz

Uncorrected and one-port cal

163

210

218

256

Two-port cal

332

425

442

487

Time Domain

3

(increase over uncorrected sweep time)

Conversions

<1

<1

4

13

Gating

<1

<1

4

17

23

IF bandwidth (Hz)

Cycle time (ms)

40,000

8

35,000

9

30,000

11

20,000

13

10,000

28

7,000

36

5,000

48

3,000

72

1,000

196

300

620

100

3853

30

8041

10

19855

Number of points Cycle time (ms)

3

4

11

4

51

5

101

6

201

9

401

16

801

29

1601

55

1. Typical performance.
2. Includes sweep time, retrace time and band-crossing

time. Analyzer display turned off with
DISPLAY:ENABLE OFF. Add 21 ms for display on.

Data for one trace (S11) measurement.

3. Option 010 only. Analyzer display turned off with

DISPLAY:ENABLE OFF. Add 21 ms for display on.