Description
Supplemental information
Line power
1
Frequency
48 Hz to 66 Hz
Voltage at 115-V setting
90 to 132 VAC; 120 VAC, nom.
Voltage at 220-V setting
198 to 264 VAC; 240 VAC, nom.
VA max
600 VA max
General environmental
RFI/EMI susceptibility
Defined by CISPR Pub. 11, Group 1, Class A, and IEC 50082-1
ESD
Minimize using static-safe work procedures and an antistatic bench mat
Dust
Minimize for optimum reliability
Operating environment
Temperature
0°C to +40°C; Instrument powers up, phase locks, and displays no error messages
within this temperature range. (Except for ‘source unleveled’ error message that may
occur at temperature extremes when power approaches limits of ALC range.)
Error-corrected temperature range
System specifications valid from 25°C ±5°C, with less than 1°C deviation from the
calibration temperature
Humidity
5 to 95% at +40°C
Altitude
0 to 4500 m (14,760 ft)
Non-operating storage environment
Temperature
-40°C to +70°C
Humidity
0 to 90% at +65°C (non-condensing)
Altitude
0 to 15,240 m (50,000 ft)
Cabinet dimensions
Exclude front and rear protrusions
Height x width x depth
267 x 425 x 426 mm, nom. (10.5 x 16.75 x 16.8 in, nom.)
Weight
Net
29 kg (64 lb), nom.
Shipping
36 kg (80 lb), nom.
1. A third-wire ground is required.