beautypg.com

Test port input, Dynamic accuracy, Continued – Atec Agilent-E8364A User Manual

Page 20

background image

Test port input

continued

Dynamic accuracy

Applies to input ports 1 and 2, accuracy of the test
port input power reading relative to the reference
input power level. Also applies to the following
conditions:

• IF bandwidth = 10 Hz
• test port powers = > –50 dBm and < 0 dBm
• magnitude dynamic accuracy = 0.02 dB +

0.001 dB/dB from reference power

• phase dynamic accuracy = 0.132° =

0.0066°/dB from reference power

20

Magnitude

0.01

0.1

1

10

-120

-110

-100

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Testport power (dBm)

Accuracy (dB)

E8362/3/4A

-10 dBm at 1 GHz

-20 dBm at 1 GHz

-30 dBm at 1 GHz

-40 dBm at 1 GHz

Phase

0.1

1

10

100

-120

-110

-100

-90

-80

-70

-60

-50

-40

-30

-20

-10

0

10

Testport power (dBm)

Accuracy (degrees)

E8362/3/4A

-10 dBm at 1 GHz

-20 dBm at 1 GHz

-30 dBm at 1 GHz

-40 dBm at 1 GHz