Measurement capabilities
Number of measurement channels
Sixteen independent measurement channels. A
measurement channel is coupled to stimulus settings
including frequency, IF bandwidth, power level,
and number of points.
Number of display windows
Up to four display windows. Each window can be
sized and re-arranged. Up to four measurement
channels can be displayed per window.
Number of traces
Up to four active traces and four memory traces
per window. Sixteen total active traces and 16
memory traces can be displayed. Measurement
traces include S-parameters, as well as relative
and absolute power measurements.
Measurement choices
S11, S21, S12, S22, A/R1, A/R2, A/B, B/R1, B/R2,
B/A, R1/A, R1/B, R1/R2, R2/A, R2/B, R2/R1, A, B,
R1, R2
Formats
Log or linear magnitude, SWR, phase, group delay,
real and imaginary, Smith chart, polar.
Data markers
Ten independent markers per trace. Reference
marker available for delta marker operation.
Marker formats include log or linear magnitude,
phase, real, imaginary, SWR, delay, R + jX, and
G + jB.
Marker functions
Marker search
Maximum value, minimum value, target, next peak,
peak right, peak left, target, and bandwidth with
user-defined target values
Marker-to functions
Set start, stop, and center to active marker stimulus
value; set reference to active marker response
value; set electrical delay to active marker phase
response value.
Trace statistics
Calculates and displays mean, standard deviation
and peak-to-peak deviation of the data trace.
Tracking
Performs new search continuously or on demand.
Source control
Measured number of points per sweep
User definable from 2 to 1601.
Sweep type
Linear, CW (single frequency), power or segment
sweep.
Segment sweep
Define up to 101 different, sub-sweep frequency
ranges in any combination of start-stop sweep
modes. Set number of points, test port power levels,
IF bandwidth, and dwell time independently for
each segment.
Sweep trigger
Set to continuous, hold, single, or group sweep
with internal or external trigger.
Power
Power slope can be set in dBm/GHz. Control the
test port signal by setting the internal attenuator of
the test set over a 60-dB range.