Atec Agilent-E8364A User Manual
Page 11
11
Extended configuration and standard power range (E8362A-Option 014)
Applies to E8362A PNA Series analyzer, 85052B (3.5 mm) calibration kit, 85131F flexible test port cable
set, and a full two-port calibration. (Specifications apply over environmental temperature of 23° ±3°C,
with less than 1°C deviation from calibration temperature.)
Description
Specification (dB)
45 MHz to 500 MHz
500 MHz to 2 GHz
2 to 10 GHz
10 to 20 GHz
Directivity
48
48
44
44
Source match
40
40
31
31
Load match
48
48
44
44
Reflection tracking
±(0.003 + 0.2/°C)
±(0.003 + 0.2/°C)
±(0.006 + 0.2/°C)
±(0.006 + 0.2/°C)
Transmission tracking
±(0.017 + 0.2/°C)
±(0.017 + 0.2/°C)
±(0.065 + 0.2/°C)
±(0.091+ 0.2/°C)
Crosstalk (reflection port to
< -85
< -100
< -110
< -100
transmission port: short
circuits at both ports;
isolation calibration
applied)
Transmission uncertainty (specifications)
Reflection uncertainty (specifications)
Magnitude
0.01
0.1
1
10
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Transmission coefficient (dB)
Uncertainty (dB)
E8362A 014 full 2-port cal using 85052B
S11 = S22 = 0
Source power = -17 dBm
■
45 MHz to 500 MHz
◆
500 MHz to 2 GHz
▲
2 GHz to 10 GHz
●
10 GHz to 20 GHz
Phase
0.1
1
10
100
-90
-80
-70
-60
-50
-40
-30
-20
-10
0
10
Transmission coefficient (dB)
Uncertainty (degrees)
E8362A 014 full 2-port cal using 85052B
S11 = S22 = 0
Source power = -17 dBm
■
45 MHz to 500 MHz
◆
500 MHz to 2 GHz
▲
2 GHz to 10 GHz
●
10 GHz to 20 GHz
Magnitude
0
0.01
0.02
0.03
0.04
0
0.2
0.4
0.6
0.8
1
Reflection coefficient (linear)
Uncertainty (linear)
E8362A 014 with 85052B
S21 = S12 = 0
Source power = -17 dBm
■
45 MHz to 500 MHz
◆
500 MHz to 2 GHz
▲
2 GHz to 10 GHz
●
10 GHz to 20 GHz
Phase
0
2
4
6
8
10
0
0.2
0.4
0.6
0.8
1
Reflection coefficient (linear)
Uncertainty (deg)
E8362A 014 with 85052B
S21 = S12 = 0
Source power = -17 dBm
■
45 MHz to 2 GHz
◆
2 GHz to 20 GHz
▲
20 GHz to 40 GHz
●
40 GHz to 50 GHz