Test port input – Atec Agilent-E8364A User Manual
Page 17
17
Test port input
Description
Specification
Supplemental information
Standard
014
UNL
UNL and 014
Test port noise floor
1
10-Hz IF bandwidth
45 to 500 MHz
2
< -89 dBm
< -89 dBm
< -89 dBm
< -89 dBm
500 MHz to 2 GHz
< -114 dBm
< -114 dBm
< -114 dBm
< -114 dBm
2 to 10 GHz
< -117 dBm
< -117 dBm
< -117 dBm
< -117 dBm
10 to 20 GHz
< -120 dBm
< -119 dBm
< -120 dBm
< -119 dBm
20 to 40 GHz
< -114 dBm
< -113 dBm
< -114 dBm
< -113 dBm
40 to 50 GHz
< -114 dBm
< -112 dBm
< -114 dBm
< -112 dBm
1-Hz IF bandwidth
45 to 500 MHz
2
< -69 dBm
< -69 dBm
< -69 dBm
< -69 dBm
500 MHz to 2 GHz
< -94 dBm
< -94 dBm
< -94 dBm
< -94 dBm
2 to 10 GHz
< -97 dBm
< -97 dBm
< -97 dBm
< -97 dBm
10 to 20 GHz
< -100 dBm
< -99 dBm
< -100 dBm
< -99 dBm
20 to 40 GHz
< -94 dBm
< -93 dBm
< -94 dBm
< -93 dBm
40 to 50 GHz
< -94 dBm
< -92 dBm
< -94 dBm
< -92 dBm
Direct receiver access input noise floor
1
10-Hz IF bandwidth
45 to 500 MHz
2
< -127 dBm
< -127 dBm
500 MHz to 2 GHz
< -133 dBm
< -133 dBm
2 to 10 GHz
< -132 dBm
< -132 dBm
10 to 20 GHz
< -134 dBm
< -134 dBm
20 to 40 GHz
< -125 dBm
< -125 dBm
40 to 50 GHz
< -123 dBm
< -123 dBm
1-Hz IF bandwidth
45 to 500 MHz
< -107 dBm
< -107 dBm
500 MHz to 2 GHz
< -113 dBm
< -113 dBm
2 to 10 GHz
< -112 dBm
< -112 dBm
10 to 20 GHz
< -114 dBm
< -114 dBm
20 to 40 GHz
< -105 dBm
< -105 dBm
40 to 50 GHz
< -103 dBm
< -103 dBm
Receiver compression level
45 MHz to 20 GHz
< 0.6 dB compression at +5 dBm
20 to 30 GHz
< 0.6 dB compression at 0 dBm
30 to 40 GHz
< 0.6 dB compression at -3 dBm
40 to 50 GHz
< 0.6 dB compression at -3 dBm
System compression level
max output power
See dynamic accuracy chart
Trace noise magnitude
45 to 500 MHz
< 0.010 dB rms
1-kHz IF bandwidth
500 MHz to 20 GHz
< 0.006 dB rms
Ratio measurement, nominal
20 to 40 GHz
< 0.006 dB rms
power at test port
40 to 50 GHz
< 0.006 dB rms
Trace noise phase
45 to 500 MHz
3
< 0.100° rms
1-kHz IF bandwidth
500 MHz to 20 GHz
< 0.060° rms
Ratio measurement, nominal
20 to 40 GHz
< 0.100° rms
power at test port
40 to 50 GHz
< 0.100° rms
1. Total average (rms) noise power calculated as
mean value of a linear magnitude trace expressed
in dBm.
2. Noise floor may be degraded by 10 dB at particular
frequencies (multiples of 5 MHz) due to spurious
receiver residuals.
3. Trace noise magnitude may be degraded to
20 mdB rms at harmonic frequencies of the first IF
(8.33 MHz) below 80 MHz.