beautypg.com

Setting the discharge volume, Operation, Operation flow – Pump Solutions Group PZiG Series User Manual

Page 16

background image

15

Operation

Before changing the mode, stop pump operation and then press the MODE key.

Setting the Discharge Volume

Operation Flow

Multiplication

times

Pump stopped

spm

Manual

MAN

Manual

MAN

Manual

MAN

Manual

STP

STP

STP

MAN

 spm

±999%

PV

(%)

%

mL/min

Set value

Set value

min

(ON)

(OFF)

Analog

ANG

Analog

ANG

 STP

Division

DIV

Pulse division

DIV

 STP

Frequency

dividing ratio

1

∼9999 times

(

✕1)

Multiplication

MUL

Pulse multiplication

MUL

 STP

Multiplication

ratio

* This setting can

be changed even
during pump
operation.

* This setting can

be changed even
during pump
operation.

* This setting can

be changed even
during pump
operation.

* The discharge volume is actually controlled every 1 spm.

The minimum controllable unit is the discharge volume x (max.
discharge volume/300) per single stroke.

* Can be changed by the keys also while pump operation is stopped or during pump operation. (Determine when the numeric value changed from blinking to light on.)

Number of

strokes setting

Changing setting values

Operation in progress

MODE

MODE

MODE

MODE

STOP

START

STOP

START

STOP

START

STOP

START

STOP

START

Count

CNT

STOP

START

Interval

INT

STOP

START

MODE

1

/1∼1/9999

1

∼9999 counts

SET

SET

SET

SET

SET

SET

SET

SET

SET

SET

SET

MODE

MODE

MODE

[Operation mode]

Number of

strokes setting

[Stop mode]

[Setup mode]

Manual

MAN

  

%

SET

STOP

START

Manual

MAN

mL/min

The discharge volume can be adjusted within the range 0.1ml / min
up to the max. discharge volume by

FUNC

FUNC

FUNC

START

Pump stops after 300 times operation.

Calibration

Determining the max. discharge volume

Determining the discharge volume

Scaling

Set point

Number of

strokes setting

SET

SET

Number of

strokes setting

SET

Proportional band

CAL

Count

CNT STP

(

✕1)

Ч1, Ч10, Ч100, Ч1000

1

∼9999 min.

1

∼9999 min.

OFF time

SET

SET

SET

ON time

Count

Interval

INT STP

(ON)(OFF)

min

Select from 0 to 9.

05-45_PZiG Series(2)English 06.8.4 9:39 AM y [ W 15