H-3531 – Xylem H-3531 FlashLite User Manual
Page 67

H-3531
SDI-12 Command and Response Protocol 5-29
5.21 Extended Read/Write Fast Mode Enable
The H-3531 FlashLite™ can be put into a Fast Mode. In this mode the atmospheric pressure and
temperature values are updated about every 3 minutes. When Fast Mode is enabled the H-3531
FlashLite™ measurement response time is 1 second. In Fast Mode the Meancount/Averaged
measurements value is not used. When the enable value is set to a ‘1' then the Fast Mode is
enabled and when it is set to a ‘0' then Fast Mode is disabled.
Command
Response
Description
“aXRFE!”
“a0011
Read Fast Mode enable
“aXWFEd!” “a0061
Write Fast Mode enable
Where:
a
is the sensor address ("0-9", "A-Z", "a-z", "*", "?").
XRFE
are upper case characters.
XWFE
are upper case characters.
d
is the new enable value(0/1)
When the user enables Fast Mode it takes 6 seconds because it updates the Atmospheric and
Temperature values until the next 3 minute update.
Example of a H-3531 Extended "Read Fast Mode Enable" command:
Command
Response
Time Values
Description
"aXRFE!"
"a0011
1 sec
1
Read Fast Mode Enable
Command
Response
Description
"aD0!"
"a+0
Fast Mode disabled
Example of a H-3531 Extended "Write Fast Mode Enable" command:
Command
Response
Time Values
Description
"aXWFE1!"
"a0061
6 sec
1
Enable Fast Mode
Command
Response
Description
"aD0!"
"a+1
Fast Mode enabled
Example of a H-3531 "aM!" command in Fast Mode:
Command
Response
Time Values
Description
"aM!"
"a0014
1 sec
4
Measure sensor in Fast Mode
Command
Response
"aD0!"
“a+AA.AA+BB.BBBB+CC.C+DD.D
See Measure command for Data response details