SDI-12 Command and Response Protocol 5-29

5.21 Extended Read/Write Fast Mode Enable
The H-3531 FlashLite™ can be put into a Fast Mode. In this mode the atmospheric pressure and
temperature values are updated about every 3 minutes. When Fast Mode is enabled the H-3531
FlashLite™ measurement response time is 1 second. In Fast Mode the Meancount/Averaged
measurements value is not used. When the enable value is set to a ‘1' then the Fast Mode is
enabled and when it is set to a ‘0' then Fast Mode is disabled.

Command

Response

Description

“aXRFE!”

“a0011

Read Fast Mode enable

“aXWFEd!” “a0061

Write Fast Mode enable

Where:

a

is the sensor address ("0-9", "A-Z", "a-z", "*", "?").

XRFE

are upper case characters.

XWFE

are upper case characters.

d

is the new enable value(0/1)

When the user enables Fast Mode it takes 6 seconds because it updates the Atmospheric and
Temperature values until the next 3 minute update.

Example of a H-3531 Extended "Read Fast Mode Enable" command:
Command

Response

Time Values

Description

"aXRFE!"

"a0011"

1 sec

1

Read Fast Mode Enable

Command

Response

Description

"aD0!"

"a+0"

Fast Mode disabled

Example of a H-3531 Extended "Write Fast Mode Enable" command:
Command

Response

Time Values

Description

"aXWFE1!"

"a0061"

6 sec

1

Enable Fast Mode

Command

Response

Description

"aD0!"

"a+1"

Fast Mode enabled

Example of a H-3531 "aM!" command in Fast Mode:
Command

Response

Time Values

Description

"aM!"

"a0014"

1 sec

4

Measure sensor in Fast Mode

Command

Response

"aD0!"

“a+AA.AA+BB.BBBB+CC.C+DD.D

See Measure command for Data response details