SDI-12 Command and Response Protocol 5-23

5.20.2 H-355 Read/Write Bubble Rate Command
This command is for reading the current bubble rate and also for editing the bubble rate.
The user can change the bubble rate from 30 to 120 bubbles/minute. Refer to H-355 Gas
Purge System manual for more details.

Command

Response

Description

“aXRBR!”

“a0051

Read Bubbler rate

“aXWBRddd!”

“a0051

Write Bubbler rate

Where:

a

is the sensor address ("0-9", "A-Z", "a-z", "*", "?").

XRBR

are upper case characters.

XWBR

are upper case characters.

ddd

is bubbles/minute value.

This command takes 005 seconds to complete and places 1 value in the data buffer. Use
the “aD0" command to collect and view the current value.

Example of a H-3531 Extended "Read Bubbler Rate" command:
Command

Response

Time

Values

Description

"aXRBR!"

"a0051"

5 sec

1

Read Bubbler Rate

Command

Response

Description

"aD0!"

"a+60"

60 bubbles/minute

Example of a H-3531 Extended "Write Bubbler Rate" command:
Command

Response

Time

Values

Description

"aXWBR80!"

"a0051"

5 sec

1

Write Bubbler Rate

Command

Response

Description

"aD0!"

"a+80"

80 bubbles/minute