5.20.2 H-355 Read/Write Bubble Rate Command
This command is for reading the current bubble rate and also for editing the bubble rate.
The user can change the bubble rate from 30 to 120 bubbles/minute. Refer to H-355 Gas
Purge System manual for more details.
Command
Response
Description
“aXRBR!”
“a0051
Read Bubbler rate
“aXWBRddd!”
“a0051
Write Bubbler rate
Where:
a
is the sensor address ("0-9", "A-Z", "a-z", "*", "?").
XRBR
are upper case characters.
XWBR
are upper case characters.
ddd
is bubbles/minute value.
This command takes 005 seconds to complete and places 1 value in the data buffer. Use
the “aD0" command to collect and view the current value.
Example of a H-3531 Extended "Read Bubbler Rate" command:
Command
Response
Time
Values
Description
"aXRBR!"
"a0051
5 sec
1
Read Bubbler Rate
Command
Response
Description
"aD0!"
"a+60
60 bubbles/minute
Example of a H-3531 Extended "Write Bubbler Rate" command:
Command
Response
Time
Values
Description
"aXWBR80!"
"a0051
5 sec
1
Write Bubbler Rate
Command
Response
Description
"aD0!"
"a+80
80 bubbles/minute