Installation 2-7

2.6 Wiring and Installation Precautions

!

The data recorder must be earth grounded.

!

The H-3531 FlashLite™ requires that the water be at ground potential. Mother
Nature takes care of this for surface and ground water.

!

Make certain the pressure port connection is tight

!

Make certain the sintered bronze vent port is not painted or covered

2.7 Programming Your SDI-12 Data Recorder
You must prepare your data recorder to receive and record the H-3531 FlashLite™ data. Since
data recorders differ widely, refer to your recorder manufacturer's directions. In general, program
the data recorder to input four values via the SDI-12 port. Usually only one or two of the
parameters are actually recorded. Your data recorder must issue an “aM!” command, then collect
the data with a “aD0" command, as explained in chapter 5. The H-3531 FlashLite™ places four
parameters in its data buffer:

a+AA.AA+BB.BBBB+CC.C+DD.D

Where:

a

= SDI-12 address 0-9, A-Z

AA.AA

= Stage (feet, inches, meters etc.)

BB.BBBB

= Pressure (PSI)

CC.C

= Temperature (C)

DD.D

= Input Supply Voltage (Volts)

2.7.1 Programming the SDI-12 Address
If more than one sensor is to be connected to the SDI-12 bus, make certain each sensor has
a different sensor address. The H-3531 FlashLite™ comes from the factory with its
address set to zero. The address can be edited using the RS-232 menu interface or using
the extended SDI-12 command. Refer to chapter 4 and 5.

2.8 Programming the H-3531 FlashLite™ Sensor
The H-3531 FlashLite™ comes from the factory with the following programmable settings:

SDI Address:

0

2

Slope:

2.3067 (feet of H 0)

Offset :

0.00

MeanCount:

8

Power_Mode:

0 = “Sleep”

0-5V_Hi:

34.6 (feet, 0-15psi)

0-5V_Lo:

0.0

4-20mA_Hi:

34.6 (feet, 0-15psi)

4-20mA_Lo:

0.0