H-3531

SDI-12 Command and Response Protocol 5-11

5.10 Extended Set Current Stage Command
The H-3531 FlashLite™ processes the pressure sensor input and computes Pressure in PSI units.
Stage is computed with a Stage = m * Pressure + b equation. During installation it is convenient
to quickly set the H-3531 FlashLite™'s Stage reading to match the current stage, elevation or
pressure of the water as determined by a staff gauge or other datum. This command causes the H-
3531 FlashLite™ to make a fresh measurement and automatically update the Offset (b) term as
needed to produce the desired Stage.

Command

Response

Description

"aXSCSddd!"

“a0061"

Set current Stage

Where: a

is the sensor address ("0-9", "A-Z", "a-z", "*", "?").

XSCS

are upper case characters.

ddd

is the new current stage value.

Example of a H-3531 Extended "Set Current Stage" command:

Command

Response

Time Values

Description

"aXSCS2.3!"

"a0061"

6 sec 1

Set the Stage to 2.3

Subsequent Command

Response

Description

"aD0"

a+1.80

The new Offset