beautypg.com

H-3531 – Xylem H-3531 FlashLite User Manual

Page 40

background image

5-2 SDI-12 Command and Response Protocol

H-3531

5.1 Measure Command
The Measure Command causes a new measurement sequence to be executed. Data values
generated in response to this command are stored in the sensor's buffer for subsequent collection
using "D" commands. The data will be retained in the sensor until another "M", " C", or "V"
command is executed.

Command

Response

Description

"aM!"

"atttn"

Initiate measurement

"aM1!"

"atttn"

Initiate measurement

Where:

a

is the sensor address ("0-9", "A-Z", "a-z", "*", "?").

M

is an upper-case ASCII character

ttt

is a three digit integer (000-999) specifying the maximum time, in seconds,
the sensor will take to complete the command and have measurement data
available in its buffer.

n

is a single digit integer (0-9) specifying the number of values that will be
placed in the data buffer. If "n" is zero (0), no data will be available using

subsequent "D" commands.

Upon completion of the measurement, a service request "a" is sent to the data recorder
indicating the sensor data is ready. The data recorder may wake the sensor with a break and
collect the data any time after the service request is received or the specified processing time has
elapsed.

Example of a H-3531 "aM!" command:
Command

Response

Time

Values

Description

"aM!"

"a0064"

6 sec

4

Make measurement

Subsequent Command

Response

"aD0"

a+AA.AA+BB.BBBB+CC.C+DD.D

Where:

AA.AA

= Stage (feet, inches, meters etc.)

BB.BBBB

= Pressure (PSI)

CC.C

= Temperature (C)

DD.D

= Power Supply Voltage (Volts)