beautypg.com

H-3531 – Xylem H-3531 FlashLite User Manual

Page 28

background image

4-4 Simple RS-232 Menu Interface

H-3531

Then the user can enter in the desired digits to the right of the decimal and press the
ENTER key to input the value. The screen will then refresh with the digits value changed.

4.3.7 O - Offset
This option is for displaying the current stage offset value and also for editing the stage
offset value manually. The following prompt appears at the bottom of the menu.

Enter Option >Enter Offset Value[ ]

Enter the desired stage offset value and press the ENTER key to input the value. The
screen will refresh and show the new offset. You must make a new measurement “M” to
see the results of the new offset.

4.3.8 U - Units
This option is for displaying the current stage units and also for editing the stage units.
The following prompt appears at the bottom of the menu.

Enter Option >Stage Units [Feet ]

Using the up and down arrows select the desired units for stage. The available stage units
are feet, meters, inches, millimeters, centimeters, psi, and user defined. After selecting the
desired units the slope will get set to the required slope to accomplish the desired units.

4.3.9 L - Slope
This option is for displaying the current stage slope and also for editing the stage slope.
The following prompt appears at the bottom of the menu.

Enter Option >Enter Slope Value[ ]

Enter the new stage slope and press the Enter key to input the value. The screen will
refresh and show the new slope. You must make a new measurement “M” to see the
results of the new stage slope.

4.3.10 C - Averaged Measurements
This option is for displaying the current Averaged Measurements/Meancount value and for
editing the value. The option also displays the estimated time that the measurement will
take due to the amount of averaging. The following prompt at the bottom of the menu.

Enter Option >Averaged Measurements [ ]

Enter the new averaged measurements/meancount and press the ENTER key. Refer to
Chapter-5 for more details on the meancount value.

4.3.11 M - Measure
Pressing the M key causes the H-3531 FlashLite™ to make a new measurement and
refresh the screen with the new measurement data. The user will see the following