5-24 SDI-12 Command and Response Protocol
H-3531
5.20.3 H-355 Read/Write Purge Pressure Command
This command is for reading the current purge pressure and also for editing the purge
pressure. The user can change the purge pressure from 15 to 80psi. Refer to H-355 Gas
Purge System manual for more details.
Command
Response
Description
“aXRPP!”
“a0051
Read Purge Pressure
“aXWPPdd!”
“a0051
Write Purge Pressure
Where:
a
is the sensor address ("0-9", "A-Z", "a-z", "*", "?").
XRPP
are upper case characters.
XWPP
are upper case characters.
dd
is the purge pressure value desired(15-80).
This command takes 005 seconds to complete and places 1 value in the data buffer. Use
the “aD0" command to collect and view the current value.
Example of a H-3531 Extended "Read Purge Pressure" command:
Command
Response
Time
Values
Description
"aXRPP!"
"a0051
5 sec
1
Read Purge Pressure
Command
Response
Description
"aD0!"
"a+60
60 PSI
Example of a H-3531 Extended "Write Purge Pressure" command:
Command
Response
Time
Values
Description
"aXWPP80!"
"a0051
5 sec
1
Write Purge Pressure
Command
Response
Description
"aD0!"
"a+80
80 PSI