5-24 SDI-12 Command and Response Protocol

H-3531

5.20.3 H-355 Read/Write Purge Pressure Command
This command is for reading the current purge pressure and also for editing the purge
pressure. The user can change the purge pressure from 15 to 80psi. Refer to H-355 Gas
Purge System manual for more details.

Command

Response

Description

“aXRPP!”

“a0051

Read Purge Pressure

“aXWPPdd!”

“a0051

Write Purge Pressure

Where:

a

is the sensor address ("0-9", "A-Z", "a-z", "*", "?").

XRPP

are upper case characters.

XWPP

are upper case characters.

dd

is the purge pressure value desired(15-80).

This command takes 005 seconds to complete and places 1 value in the data buffer. Use
the “aD0" command to collect and view the current value.

Example of a H-3531 Extended "Read Purge Pressure" command:
Command

Response

Time

Values

Description

"aXRPP!"

"a0051"

5 sec

1

Read Purge Pressure

Command

Response

Description

"aD0!"

"a+60"

60 PSI

Example of a H-3531 Extended "Write Purge Pressure" command:
Command

Response

Time

Values

Description

"aXWPP80!"

"a0051"

5 sec

1

Write Purge Pressure

Command

Response

Description

"aD0!"

"a+80"

80 PSI