1.2.2 IDS adapter for analog pH sensors
With the aid of an IDS adapter, you can also operate analog sensors
on the MultiLab 4010-2. The combination of the IDS adapter and analog sensor
behaves like an IDS sensor.
The MultiLab 4010-2 has a recess where the IDS adapter (ADA 94/IDS DIN or
ADA 94/IDS BNC), which is available as an accessory, can be permanently
mounted.
The IDS adapter replaces a digital input (channel 2) with a connector for an
analog pH sensor (DIN or BNC plug) and a temperature sensor.
1.2.3 Automatic sensor recognition
The automatic sensor recognition for IDS sensors allows
to operate an IDS sensor with different meters without recalibrating
to operate different IDS sensors at one meter without recalibration
to assign measurement data to an IDS sensor
– Measurement datasets are always stored and output with the sensor
name and sensor series number.
to assign calibration data to an IDS sensor
– Calibration data and calibration history are always stored and output with
the sensor name and sensor series number.
to activate the correct cell constant for conductivity sensors automatically
to hide menus automatically that do not concern this sensor
To be able to use the automatic sensor recognition, a meter that supports the
automatic sensor recognition (e.g. MultiLab 4010-2) and a digital IDS sensor
are required.
In digital IDS sensors, sensor data are stored that clearly identify the sensor.
The sensor data are automatically taken over by the meter.