4 cond measurement settings, 1 settings for ids conductivity sensors, Cond measurement settings – Xylem MultiLab 4010-2 User Manual
Page 68: Settings for ids conductivity sensors, Ee section 9.4.1 s, Tings, Conductivity, Sensors
Settings
MultiLab 4010-2
68
ba76142e01
05/2013
D.O. mg/L readings are unaffected by the selection of Saturation local.
9.4
Cond measurement settings
9.4.1 Settings for IDS conductivity sensors
The settings are made in the menu for the measured parameter, conductivity.
To open the settings, display the required measured parameter in the mea-
sured value display and press the
switch to the measured value display with
The possible settings are individually displayed for each sensor. Below the set-
ting menu is displayed for two IDS sensors (
4310, 4320).
Default settings are printed in bold.
Setting menu of
4310
Menu item
Possible
setting
Explanation
Calibration /Calibra-
tion record
-
Displays the calibration record of the
last calibration.
Calibration / Calibra-
tion data storage /
Display
-
Displays the last calibration records
(max. 10)
Calibration / Calibra-
tion data storage /
Output to USB flash
drive or printer
-
Outputs the stored calibration data to
the USB-A interface (USB memory
device/USB printer)
Calibration /
Calibration data stor-
age / Output to
RS232/USB
-
Outputs the stored calibration data to
the USB-B interface (PC)
Calibration /Calibra-
tion interval
1 ... 150 ...
999 d
Calibration interval for the IDS con-
ductivity sensor (in days).
The meter reminds you to calibrate
regularly by the flashing sensor sym-
bol in the measuring screen.
Type
Used measuring cell
Cal
Measuring cells whose cell constant
is determined by calibration in the KCl
control standard solution.
Calibration range:
0.450 to 0.500 cm
-1
The currently valid cell constant is dis-
played in the status line.
man
Freely adjustable cell constant in the
range 0.450 ... 0.500 cm
-1
.
Man. cell const.
0.450 ...
0.475 ...
0.500 cm
-1
Display and setting options for the cell
constant.