MultiLab 4010-2

Dissolved oxygen (D.O.)

ba76142e01

05/2013

51

3.

Place the D.O. sensor in a solution that does not contain any dissolved
oxygen.

A solution not containing any dissolved oxygen can be prepared by
dissolving approx. 8 to 10 g sodium sulfite (Na

2

SO

3

) in 500 ml tap-

water. Carefully mix the solution. It may take up to 60 minutes until
the solution is free of oxygen.

4.

Start the calibration with .
Cal (calibration in water vapor-saturated air) or Comp (calibration via
Comparison meas. according to the Winkler method) is displayed.
The calibration point 1 (0%) is displayed.

5.

Start the measurement with .
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter flashes.

6.

Wait for the AutoRead measurement to be completed (audio signal).
The display for the second calibration point is displayed.