2 enter sensor cap coefficients, 3 saturation local, Enter sen – Xylem MultiLab 4010-2 User Manual
Page 67: Satur, Section 9.3.3 s, Aturation local
MultiLab 4010-2
Settings
ba76142e01
05/2013
67
9.3.2 Enter Sensor Cap coefficients
9.3.3 Saturation local
Irrespective of the height or air pressure, the calibration value is set to 100%.
Saturation local is ideal for EU compliance.
When Saturation local is enabled, an L will appear next to D.O.% on the
display.
Calibration /
Sensor Cap coeffici
ents
K1
...
K5
KC
Enter the cap coefficients here after
exchanging the sensor cap.
For details, see section 9.3.2 E
Salinity
0.0 ... 70.0
Salinity or salinity equivalent for the
salt content correction.
This menu item is only available if the
automatic salt content correction is
switched off and the manual salt con-
tent correction is switched on.
Resolution
0.1
1
Set a high or low resolution.
The setting of the resolution is sepa-
rately stored for each measured
parameter.
Saturation local
on
off
Saturation local is a procedure that
takes the local air pressure into
account for each saturation measure-
ment.
For details, see section 9.3.3
S
Reset
-
Resets all sensor settings to the deliv-
ery condition (see section 9.6.1 R
The values of the coefficients are provided with the sensor cap.
1.
Change the digit of the highlighted position with <><>.
2.
Go to the next position with <><>.
3.
Confirm with
Menu item
Possible
setting
Explanation