beautypg.com

Xylem MultiLab 4010-2 User Manual

Page 42

background image

pH value

MultiLab 4010-2

42

ba76142e01

05/2013

sor evaluation.

Carrying out a QSC

control calibration

A QSC control calibration can, e.g. be useful if the sensor evaluation noticeably
changed (after some normal calibrations).

You can carry out QSC control calibrations at greater intervals than normal cali-
brations.

1 QSC scale

The double arrow on the QSC scale indicates the current sensor eval-
uation

03.04.2013 08:00

Info

1

1.

Open the menu for measurement settings with .

2.

In the QSC menu, select Control calibration with <><>.
The calibration display appears. AutoCal QSC-Kit is displayed as the buf-
fer.
Exclusively use the QSC-Kit for the QSC calibration. If you use other buf-
fers, you will have no valid QSC control calibration.

3.

Follow the user guide.
The calibration is carried out like a normal three-point calibration. As soon
as the three-point calibration has been successfully carried out you can
decide whether to accept or discard the calibration as the QSC control
calibration.