6 calibration data, Calibration data – Xylem MultiLab 4010-2 User Manual
Page 53
MultiLab 4010-2
Dissolved oxygen (D.O.)
ba76142e01
05/2013
53
7.2.6 Calibration data
Displaying the
calibration record
The calibration record of the last calibration is to be found under the menu item,
Calibration / Calibration record. To open it in the measured value display, press
the
_
> key.
The calibration records of the last 10 calibrations are to be found in the menu,
Calibration / Calibration data storage / Display. To open the Calibration menu,
press the
Calibration
evaluation
After calibration, the meter automatically evaluates the current status of the
calibration. The evaluation appears on the display and in the calibration record.
The calibration record is automatically transmitted to the interface
after calibrating.
Menu item
Setting/func-
tion
Explanation
Calibration /
Calibration data
storage /Display
-
Displays the calibration records.
Further options:
Scroll through the calibration
records with <><>.
Using
played calibration record to
the USB-B (PC) interface or
the USB-A (USB printer)
interface .
Using
_
>, output all
calibration records to the
USB-B (PC) interface or the
USB-A (USB printer) inter-
face .
Quit the display with
Switch directly to the mea-
sured value display with
Calibration / Cali-
bration data stor-
age / Output to
USB flash drive or
printer
-
Outputs the stored calibration
data to the USB-A interface
(USB memory device/USB
printer)
Calibration /
Calibration data
storage / Output to
RS232/USB
-
Outputs the stored calibration
data to the USB-B interface (PC)