9.1.3 Calibration interval
The calibration evaluation is displayed as a sensor symbol.
After the QSC function has been enabled the sensor symbol is replaced by the
QSC scale (see section 5.2.8 QSC
40).
After the specified calibration interval has expired the sensor symbol or the
QSC scale flashes. It is still possible to measure.
20
Reagecon TEC *
2.000
4.010
7.000
10.000
25 °C
21
Reagecon 20 *
2.000
4.000
7.000
10.000
13.000
20 °C
22
Reagecon 25 *
2.000
4.000
7.000
10.000
13.000
25 °C
23
Chemsolute *
2.000
4.000
7.000
10.000
20 °C
24
USABlueBook
*
4.000
7.000
10.000
25 °C
25
YSI *
4.000
7.000
10.000
25 °C
The buffers are selected in the menu, pH /
/ Buffer (see section 9.1.1 S
61).
To ensure the high measuring accuracy of the measuring system,
calibrate after the calibration interval has expired.