4 calibration data, Calibration data – Xylem MultiLab 4010-2 User Manual
Page 59
MultiLab 4010-2
Conductivity
ba76142e01
05/2013
59
8.3.4 Calibration data
The calibration data can be displayed and then output to the interface.
Displaying the cali-
bration record
The calibration record of the last calibration is to be found under the menu item,
Calibration / Calibration record. To open it in the measured value display, press
the
_
> key.
The calibration records of the last 10 calibrations are to be found in the menu,
Calibration / Calibration data storage / Display. To open the Calibration menu,
press the
4.
Immerse the conductivity sensor in the control standard solution,
0.01 mol/l KCI.
5.
Start the measurement with
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.
6.
Wait for the end of the measurement with stability control
or
take over the calibrated value with
The calibration record is displayed and output to the interface.
7.
Switch to the measured value display with
The calibration record is automatically transmitted to the interface
after calibrating.
Menu item
Setting/
function
Explanation
Calibration /
Calibration data
storage /Display
-
Displays the calibration records.
Further options:
Scroll through the calibration
records with <><>.
Using
played calibration record to the
USB-B (PC) interface or the
USB-A (USB printer) interface .
Using
_
>, output all cali-
bration records to the USB-B
(PC) interface or the USB-A
(USB printer) interface .
Quit the display with
Switch directly to the measured
value display with