8.3.4 Calibration data

The calibration data can be displayed and then output to the interface.

Displaying the cali-

bration record

The calibration record of the last calibration is to be found under the menu item,
Calibration / Calibration record. To open it in the measured value display, press
the

_

> key.

The calibration records of the last 10 calibrations are to be found in the menu,
Calibration / Calibration data storage / Display. To open the Calibration menu,
press the key in the measured value display.

4.

Immerse the conductivity sensor in the control standard solution,
0.01 mol/l KCI.

5.

Start the measurement with .
The measured value is checked for stability (stability control).
The [AR] status indicator is displayed. The measured parameter
flashes.

6.

Wait for the end of the measurement with stability control
or
take over the calibrated value with .
The calibration record is displayed and output to the interface.

7.

Switch to the measured value display with .

The calibration record is automatically transmitted to the interface
after calibrating.

Menu item

Setting/
function

Explanation

Calibration /
Calibration data
storage
/Display

-

Displays the calibration records.

Further options:

Scroll through the calibration
records with <><>.

Using , output the dis-
played calibration record to the
USB-B (PC) interface or the
USB-A (USB printer) interface .

Using

_

>, output all cali-

bration records to the USB-B
(PC) interface or the USB-A
(USB printer) interface .

Quit the display with .

Switch directly to the measured
value display with .