Settings
9.1
pH measurement settings
9.1.1 Settings for pH measurements
The settings are made in the menu for calibration and measurement settings of
the pH/ORP measurement. To open the settings, display the required mea-
sured parameter in the measured value display and press the
After completing the settings, switch to the measured value display with
Default settings are printed in bold.
Menu item
Possible
setting
Explanation
Calibration /Calibration
record
-
Displays the calibration record of the last calibration.
Calibration / Calibration
data storage / Display
-
Displays the last calibration records (max. 10)
Calibration /
Calibration data storage
/ Output to RS232/USB
-
Outputs the stored calibration data to the USB-B interface
(PC)
Calibration / Calibration
data storage / Output to
USB flash drive or
printer
-
Outputs the stored calibration data to the USB-A interface
(USB memory device/USB printer)
Calibration /Buffer
YSI
NIST/DIN
ConCal
...
Buffer sets to be used for pH calibration.
More buffers and details: see section 9.1.2 B
Calibration /Single-point
calibration
yes
no
Quick calibration with 1 buffer
Calibration /Calibration
interval
1 ... 7 ... 999
d
Calibration interval for the IDS pH sensor (in days).
The meter reminds you to calibrate regularly by the flash-
ing sensor symbol in the measuring screen.
Calibration /Unit for
slope
mV/pH
%
Unit of the slope.
The % display refers to the Nernst slope of
-59.2 mV/pH (100 x determined slope/Nernst slope).
QSC /First calibration
-
Starts the initial calibration with QSC buffers.
This menu item is only available as long as no initial cali-
bration was carried out with the connected IDS sensor.
QSC /Record of first cal-
ibration
-
Displays the calibration record of the QSC initial calibra-
tion.
QSC /Control calibra-
tion
-
Starts the control calibration with QSC buffers.
This menu item is only available if an initial calibration was
carried out with the connected IDS sensor.