Xylem MultiLab 4010-2 User Manual
Page 74
Settings
MultiLab 4010-2
74
ba76142e01
05/2013
Dissolved oxygen
The following settings for D.O. measurements are reset to the default settings
with the Reset function:
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the
key.
Conductivity
The following settings for conductivity measurements are reset to the default
settings with the Reset function:
The sensor settings are reset under the Reset menu item in the menu for cali-
bration and measurement settings. To open the settings, display the required
measured parameter in the measured value display and press the
key.
Setting
Default settings
Calibration interval
180 d
Check interval
60 d
Measured parameter
D.O. concentration
Relative slope (S
Rel
)
1.00
Salinity (value)
0.0
Salinity (function)
Off
Setting
Default settings
Calibration interval
150 d
Measured parameter
ϰ
Cell constant (c)
Depending on the connected mea-
suring cell:
0.475 cm
-1
(calibrated)
0.475 cm
-1
(set)
0.100 cm
-1
Temperature compensation
nLF
Reference temperature
25 °C
Temperature coefficient (TC) of the
linear temperature compensation
2.000 %/K
Multiplier for TDS
1.00