Calculating drift per sensor, Appendix d: calculating drift per sensor, Appendix d – Ocean Optics NeoFox User Manual
Page 74

Appendix D
Calculating Drift Per Sensor
Drift differs depending on the sensor formulation (FOXY, FOSPOR or HIOXY). See the table and
example below.
Stability (Continuous LED):
FOXY Formulation
FOSPOR Formulation
HIOXY Formulation
Lifetime Stability
(Tau):
0.0006 usec/ HR
0.003 usec/ HR
0.0002 usec/HR
Oxygen Stability %:
0.01% HR
0.005% HR
0.007% HR
Example: Calculate drift for a FOXY sensor probe while measuring a sample every 2.5 hours for 100
hours with the LED being lit for 1 minute:
New drift = reported drift * duty cycle on / duty cycle off
New drift = 0.01%/hr * (1 min / (2.5 hr*60 min/hr)) 2.5 * 60 converts hours into minutes,
so as to cancel out the dimensions
New drift = 0.01%/hr *(1 min/150 min)
New drift = 0.01%/hr * (.0066)
New drift = 0.000066 %/ hr
To run the experiment for 100 hours:
New drift = 0.000066%/hr * 100 hrs
New drift = 0.006%
013-20000-009-02-201203
64
- Apex 785 Raman (1 page)
- STS-UV (2 pages)
- TR2 Engineering Note (4 pages)
- SAD500 Communications and Control (19 pages)
- Red Tide USB650 Install (26 pages)
- Fiber Optic Termination Kit (6 pages)
- Transmissive pH Probe (10 pages)
- Remora (42 pages)
- PlasCalc (59 pages)
- Correcting Device Driver Issues (8 pages)
- ecoVis Krypton Light Source (16 pages)
- LPC-500CM (28 pages)
- HPX-2000 (24 pages)
- ADC1000-USB (27 pages)
- Torus Operating Instructions (30 pages)
- ADC2000-PCI (13 pages)
- Sensors for Real-Time Analysis (2 pages)
- IDRaman reader (2 pages)
- DH-2000-CAL (30 pages)
- QE65 Pro (32 pages)
- Collimating Lenses (2 pages)
- D1000 (2 pages)
- IDRaman mini (2 pages)
- HR2000CG-UV-NIR (42 pages)
- Cool Red (1 page)
- HL-2000 (20 pages)
- XE-1 Xenon (6 pages)
- USB-ADP Serial Adapters (3 pages)
- CHEM2000 (36 pages)
- Deuterium-Halogen Calibration Light Source (19 pages)
- NeoFox Engineering Note (30 pages)
- QE Pro (2 pages)
- OOIBase32 (140 pages)
- AR-1 Argon (6 pages)
- pH Sensor Patches, Probes and Cuvettes (36 pages)
- SpecLine Offline Spectroscopy (60 pages)
- HL-2000-HP-232 (26 pages)
- External Triggering Options Instructions for Spectrometers with Firmware Version 3.0 and Above (16 pages)
- Breakout Box (10 pages)
- USB-ISS-UV_VIS (4 pages)
- ISS-UV_VIS (6 pages)
- OOIColor (14 pages)
- LS-1 Series (12 pages)
- Apex Install (24 pages)
- DH-2000 (34 pages)