Calibrating the probe, Multipoint calibration – Ocean Optics NeoFox User Manual

Page 35

background image

3: Using NeoFox with NeoFox Viewer Software

013-20000-009-02-201203

25

2. Click

the

button in the Gain field.

3. Slide the APD Voltage selector to the desired level.

4. Click

the

button in the Gain field.

Calibrating the Probe

When you access the software for the first time, the Oxygen field will be blank. To see an oxygen value,
you must first enter calibration information for the probe. This calibration information is unique for each
oxygen sensor probe itself, but is independent of the NeoFox hardware unit. Therefore, a new calibration
must be loaded each time a new sensor is added or before first use.

You have the option of purchasing a factory calibration for your Oxygen Sensor Probe (FOXY-CAL,
HIOXY-CAL, FOSPOR-CAL). If you have a factory calibration, follow the instructions for

Multipoint

Calibration

. If you do not have a factory calibration, follow the instructions for a

Two-Point Calibration

.

Multipoint Calibration

Follow these instructions if you purchased a factory calibration file for your sensor. This file can be found
on the NeoFox CD. To retrieve it, select Calibration from the Options menu.