Calibrating the probe, Multipoint calibration – Ocean Optics NeoFox User Manual
Page 35
3: Using NeoFox with NeoFox Viewer Software
013-20000-009-02-201203
25
2. Click
the
button in the Gain field.
3. Slide the APD Voltage selector to the desired level.
4. Click
the
button in the Gain field.
Calibrating the Probe
When you access the software for the first time, the Oxygen field will be blank. To see an oxygen value,
you must first enter calibration information for the probe. This calibration information is unique for each
oxygen sensor probe itself, but is independent of the NeoFox hardware unit. Therefore, a new calibration
must be loaded each time a new sensor is added or before first use.
You have the option of purchasing a factory calibration for your Oxygen Sensor Probe (FOXY-CAL,
HIOXY-CAL, FOSPOR-CAL). If you have a factory calibration, follow the instructions for
. If you do not have a factory calibration, follow the instructions for a
Multipoint Calibration
Follow these instructions if you purchased a factory calibration file for your sensor. This file can be found
on the NeoFox CD. To retrieve it, select Calibration from the Options menu.