Analog output parameters list, Analog output specifications – Ocean Optics NeoFox User Manual
Page 69
C: Analog Output
013-20000-009-02-201203
59
Analog Output Parameters List
NeoFox Viewer Parameter
Potential Values
Voltage Output Source
0 – Disable analog output
1 – Oxygen
2 – Temperature (not supported)
3 – Pressure (not supported)
4 – Tau
5 – Alternate O
2
Units
Current Output Source
0 – Disable analog output
1 – Oxygen
2 – Temperature (not supported)
3 – Pressure (not supported)
4 – Tau
5 – Alternate O
2
Units
Analog Voltage Lower Bound (or Voltage
Limit Minimum)
- This floating point value represents
the “lower bound” of the represented range of the input
data. Must be < UBOUND and ≥ 0.
Analog Voltage Upper Bound (or Voltage
Limit Maximum)
- This floating point value represents
the “upper bound” of the represented range of the input
data. Must be > LBOUND.
Analog Current Lower Bound (or Current
Limit Minimum)
- This floating point value represents
the “lower bound” of the represented range of the input
data. Must be < UBOUND and ≥ 0.
Analog Current Upper Bound (or Current
Limit Maximum)
- This floating point value represents
the “upper bound” of the represented range of the input
data. Must be > L
Analog Output Specifications
Name Min
Typ
Max
Conditions
4-20 mA Loop Voltage
11 V
--
24V
R
Loop
= 2 Ω
4-20 mA Output Accuracy
--
< 160 μA --
25°C,
V
Loop
= 12V, R
Loop
= 2 Ω
0-5V Output Accuracy
--
< 50mV
--
25°C
0-5V Output Resistance
--
--
250 Ω --