Two-point calibration – Ocean Optics NeoFox User Manual

Page 37

background image

3: Using NeoFox with NeoFox Viewer Software

013-20000-009-02-201203

27

8. Click

on

button on the bottom right corner of the screen. The Computed

Calibration Constants table fills in on the bottom left corner of the screen.

9. Click

the

button to use the coefficients for oxygen calculation. A Download

Confirmation screen appears.

10. Click Yes to confirm the download and continue.

11. Proceed to the

Single Point Reset

.

Two-Point Calibration

If you want to do your own calibration instead of using a factory calibration, follow these Two-Point
Calibration instructions.

Procedure

1. From the Main screen, select Options | Calibration.

2. Click

the

Setup tab. The Setup screen appears.